Study on Accuracy of Scanned Impressions of Edentulous Mandibular Implants and Conventional Impressions using Digital Wings Scanning Body

HUANG Jing1, LU Qinyi2, XU Peng2, LIANG Yuan2*   

  1. 1. Chongqing Jingmei Medical Technology Co.,Ltd., Chongqing 400010, China;
    2. Chongqing Medical University Affiliated Stomatological Hospital, Chongqing 400015, China
  • Received:2024-12-05 Published:2025-10-23

摘要: 目的: 探究Digital Wings新型扫描体对无牙颌种植口内扫描印模技术精度的影响,分析其在数字化口内扫描组中的表现,以及与传统印模技术的精度的差异,为临床应用提供参考。方法: 使用最佳拟合法和绝对线性偏差法得到传统取模组、口内扫描组和标准模型组的种植体位置的准确性和精度有效值(root mean square, RMS)。结果: 传统组、普通扫描体组、新型CAD扫描体组和Digital Wings扫描体组准确性的均方根RMS中位数(四分位间距)分别为39.3(29.9~50.5) μm、135.1(50.8~260.5) μm、84.3(53.2~212.5) μm、81.3(33.8~133.2) μm。Digital Wings扫描体组和传统取模组比较无显著性差异(P>0.05)。精度的RMS均数分别为35.3(12.4~98.5) μm、155.4(33.2~329.9) μm、116.4(4.4~349.8) μm、54.0(13.4~124.6) μm。Digital Wings扫描体组和新型CAD扫描体组与传统取模组比较无显著差异(P>0.05)。结论: Digital Wings扫描体组的精度和传统组的精度比较无显著差异。扫描体的改进,如扫描体外形的改变使扫描体的精度逐渐增加。更多关于无牙颌种植扫描体体外研究和临床研究需要开展。

关键词: 精度, 准确性, Digital Wings, 扫描体, 无牙颌

Abstract: Objective: To investigate the effect of Digital Wings scanning body on the accuracy of oral scanning impression technique for edentulous maxillary implants, analyze its performance in several intraoral scanning groups, and compare its accuracy with that of the traditional impression technique. Methods: Root mean square (RMS) values for the trueness and precision of implant positions were obtained for conventional impression-taking group, intraoral scanning group, and standard model group using the best-fit method and the absolute linear deviation method. Results: RMS means of trueness for the conventional group, the normal scanner group, the new CAD scanner group, and the Digital Wings scanner group were 39.3 (29.9-50.5) μm, 135.1 (50.8-260.5) μm, 84.3 (53.2-212.5) μm, and 81.3 (33.8-133.2) μm, respectively. No significant difference was observed between the Digital Wings scanner group and the conventional mode-taking group (P>0.05). The RMS means of precision were 35.3 (12.4-98.5) μm, 155.4 (33.2-329.9) μm, 116.4 (4.4-349.8) μm, and 54.0 (13.4-124.6) μm, respectively. Additionally, no significant difference was found between the Digital Wings scanner group, the new CAD scanner group, and the traditional mode-taking group (P>0.05). Conclusion: There was no statistically significant difference in accuracy between the Digital Wings scanner group and the conventional group. Improvements in scan body design, such as changes in the shape of the scan body, have led to a gradual increase in scanning accuracy. However, more in vitro and clinical studies on edentulous implant scan body are needed.

Key words: precision, trueness, Digital Wings, scan body, edentulous