Journal of Oral Science Research ›› 2016, Vol. 32 ›› Issue (11): 1196-1199.DOI: 10.13701/j.cnki.kqyxyj.2016.11.021

Previous Articles     Next Articles

Accuracy of CAD/CAM Template for Implant Site Preparation

WANG Jin-sheng1, XU Liang-wei2, JIANG Chen-bo1, XU Ban2*, SHAO Li-min3, ZHANG Fei-fei4   

  1. 1. Department of Stomatology, Lanxi Peoples'Hospital, Jinhua 321199, China;
    2. Zhejiang University of Technology, Hangzhou 310000, China;
    3. Shao Xiaowei Traditional Chinese Medicine Hospital of Lanxi City, Jinhua 321102, China;
    4. Department of Stomatology, The Second Affiliated Hospital, College of Medicine, Zhejiang University, Hangzhou 310009, China
  • Received:2016-04-25 Online:2016-11-25 Published:2016-11-25

Abstract: Objective: To evaluate the accuracy of the CAD/CAM implant template based on CBCT data for dental implants preparation. Methods: Jaws from 6 patients were scanned using cone beam computed tomography (CBCT). The CT data was used to fabricated CAD/CAM implant templates. Twenty-four implants were inserted in edentulous patients using CAD/CAM implant templates. All patients had CBCT scanning again after operation. Preoperative and postoperative computed tomographic images were compared, and the deviations between actual and planned implants were measured. Results: With the help of CAD/CAM implant templates, the neck deviation of the implant was 1.03±0.55mm, apical deviation was 1.19±0.56mm, and axis deviation was 3.12±2.64°. Conclusion: The CAD/CAM implant template can accurately transfer the preoperative design, reduce the risk of operation, and is worth for clinical use.

Key words: CAD/CAM, Implant template, Accuracy, CBCT